1WMPD4001402B
FXi-11 Large Breeze Break
1. Features
The FXi-11 can be installed on the FX-i series balances.
Using the FXi-11 will protect the balance from the effects of a breeze and will provide more accurate
weighing results.
An anti-static treatment has been applied to the breeze break components, therefore the weighing error
caused by static electricity should not be a problem.
2. Unpacking
Large breeze break
3. Installing the large breeze break
1. Remove the weighing pan and the pan
support from the balance.
2. Remove the two screws that secure the dust
plate. Remove the dust plate from the balance.
3.
Confirm that the “UPPER SIDE” label is visible (up).
Using the two screws supplied, install the breeze
break lower plate on the balance.
CAUTION
When securing, confirm
that the spacing
between the inside of
breeze break lower
plate and the outside of
the pan support boss is
balanced.
CAUTION - When installing, use care not to apply excessive load to the sensor mechanism
(circumference of the pan support boss) of the balance.
- When securing the breeze break lower plate, use the accessory screws provided with
the FXi-11. (Keep the original dust plate and screws for later reinstallation.)
4. Pull out the handles located on the each side of
the breeze break and install the large breeze
break on the balance.
5. Push in the handles from each side.
6. Install the pan support on the pan support boss and install the weighing pan on the pan support.
CAUTION - When transporting the balance, do not hold the
breeze break. Hold the balance.
- Do not wipe the transparent plastic with
solvents such as thinner, benzene or
alcohol. It may cause it to turn white.
- An anti-static treatment has been applied to the
transparent plastic, therefore do not use solvents
such as alcohol to wipe the breeze break.
- When cleaning the breeze break; wipe with a dry,
lint free cloth or a lint free cloth that is moistened
with warm water and a mild detergent.
- If the anti-static effect decreases, apply an
anti-static additive to the transparent plastic.
Materials used (Material of the breeze break)
Transparent plastic: An anti-static treated
polycarbonate
Slide pin: Polyacetal Other plastic: Electro-conductive ABS
Screws: steel Other metal: SUS
Breeze break
lower plate
Screws for securing
the breeze break
lower plate (2 pcs)
Driver (+)
Weighing
pan
Pan
support
Screws for
securing the
dust plate
(2 pcs)
Dust plate
Balance
1
2
3
breeze break
lower plate
Pan support
boss
Screws for securing
the breeze break
lower plate (2 pcs)
Breeze break lower
plate
Pan support
boss
Balance
4
4
5
5
Top slide screws
Opening direction
of the top door
depends on which
of the screws is
removed.
Opening
direction of
the top door
F
X
i
-11
Large breeze
break
Pull out
the handles
Balance
Breeze break
lower plate
Confirm that the
handle’s hooks
(each side) are
under the
breeze break
lower plate.
Push in the
handles
Opening
direction of
the top door
Top door