7-4-4 RsVisco Format
Used with the graphing program RsVisco. Measurement data are output in comma separated format.
When a comma is selected as the decimal point by "pnt 1", a semicolon ";" is used as a data separator.
When a measurement is started using RsVisco, the viscometer automatically selects this format.
Measurement data are output in the order of viscosity, viscosity unit, temperature and temperature
unit.
The output data are 25 characters long excluding the terminator
With RsVisco format selected, the viscosity value and the temperature are output using the internal resolution.
The relation between the measuring unit and the internal resolution is as follows:
Viscosity Temperature
Model
mPas Pas cP P
° °
SV-10 0.01 0.0001 0.01 0.0001 Internal
resolution
SV-100
- 0.01 - 0.1
0.01 0.01
SV-10 output format example
Viscosity
/
Temper-
ature
Display Output format example Remarks
LLLLL mPas
+00000.00,mPa s,+025.67,C
Zeroes are output for below measuring
range error.
0.30 mPas
+00000.30,mPa s,+025.67,C
10.0 mPas
+00010.00,mPa s,+025.67,C
100 mPas
+00100.00,mPa s,+025.67,C
1.00 Pas
+01000.00,mPa s,+025.67,C
For 1000 mPas or greater, the displayed unit
is Pas, but the output unit remains mPas.
mPas
/
°C
H Pas
+12000.00,mPa
s,+025.67,C
12000 is output for above measuring
range error.
L Pas
+000.0000, Pa s,+051.23,F
Zeroes are output for below measuring
range error.
0.0003 Pas
+000.0003,
Pa s,+051.23,F
0.0100 Pas
+000.0100,
Pa s,+051.23,F
0.100 Pas
+000.1000, Pa s,+051.23,F
1.00 Pas
+001.0000, Pa s,+051.23,F
Pas
/
°F
H Pas
+012.0000,
Pa ,+051.23,F
12 is output for above measuring range error.
L cP
+000.0000, cP ,+025.67,C
Zeroes are output for below measuring
range error.
0.30 cP
+00000.30, cP ,+025.67,C
10.0 cP
+00010.00, cP ,+025.67,C
100 cP
+00100.00, cP ,+025.67,C
10.0 P
+01000.00, cP ,+025.67,C
For 1000 cP or greater, the displayed
unit is P, but the output unit remains cP.
cP
/
°C
H P
+12000.00,
cP ,+025.67,C
12000 is output for above measuring
range error.
L P
+000.0000,
P ,+051.23,F
Zeroes are output for below measuring
range error.
0.0030 P
+000.0030,
P ,+051.23,F
0.100 P
+000.1000,
P ,+051.23,F
1.00 P
+001.0000, P ,+051.23,F
10.0 P
+010.0000, P ,+051.23,F
P
/
°F
H P
+120.0000,
P ,+051.23,F
120 is output for above measuring range error.