FZ-i / FX-i SERIES FZ-120i / FZ-200i / FZ-300i FZ-1200i / FZ-2000i / FZ-3000i FX-120i / FX-200i / FX-300i FX-1200i / FX-2000i / FX-3000i Precision B
8 This balance uses a strong magnet as part of the balance assembly, so please use caution when weighing magnetic materials such as iron. If there
93-5 Display Symbols and Key Operation Key operation Key operation affects how the balance functions. The basic key operations are: “Press and re
104. WEIGHING UNITS 4-1 Units With the FZ-i / FX-i series balance, the following weighing units and weighing modes are available: A unit or m
11The tables below indicate the weighing capacity and the minimum display for each unit, depending on the balance model. Capacity Unit FZ-120i FX-12
124-2 Storing Units The units or modes can be selected and stored in the function table. The sequence of displaying the units or modes can be arrange
135. WEIGHING 5-1 Basic Operation (Gram Mode) 1 Place a container on the weighing pan, if necessary. Press the RE-ZERO key to cancel the weight
145-2 Counting Mode (PCS) This is the mode to determine the number of objects in a sample based on the standard sample unit mass. Unit mass means th
15 If the balance judges that the mass of the samples is too light to aquire accurate weighing, it displays an error requiring the addition of more s
165-3 Percent Mode (%) This is the mode to display the weight value in percentage compared to a 100% reference mass and is used for target weighing
175-4 Statistical Calculation Mode The statistical calculation mode statistically calculates the weight data, and displays or outputs the results. T
© 2008 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any l
18Selecting the statistical items to output 5 Press the SAMPLE key to display 5taf 0 . 6 Press the RE-ZERO key to select the output items.
19 1 Press the RE-ZERO key to set the display to zero. 2 Place the sample on the weighing pan and wait for the stabilization indicator to turn on
20Deleting the latest data When the wrong data is entered, it can be deleted and excluded from statistical calculation. Only the latest data can be
215-5 Statistical Calculation Mode (Example of Use) Here, as an example of use of the statistical calculation mode, mixing of the multiple formulae
225-5-2 Using The Statistical Calculation Mode 1 Press the RE-ZERO key to set the display to zero. 2 Place a container on the weighing pan. Pr
236. RESPONSE ADJUSTMENT This function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and vibration at the
247. CALIBRATION 7-1 Calibration Mode The FX-i series balance has the following two modes. Calibration using the internal mass (One-touch calibra
257-2 Calibration Using the Internal mass (One-Touch Calibration, only for FZ-i series) This function calibrates the balance using the internal mas
267-3 Correcting the Internal Mass Value (Only for FZ-i series) The FZ-i series balance can correct the internal mass value by basing on an external
277-4 Calibration Using an External Weight This function calibrates the balance using an external weight. Operation 1 Connect the AC adapter and
1CONTENTS Basic Operation 1. INTRODUCTION ...
28 5 Confirm that there is nothing on the pan and press the PRINT key. The balance measures the zero point. Do not allow vibration or drafts to aff
297-5 Calibration Test Using an External Weight This function tests the balance weighing accuracy using an external weight and outputs the result. T
30 5 Confirm that there is nothing on the pan and press the PRINT key. The balance measures the zero point and displays the measured value. Do not
318. FUNCTION SWITCH AND INITIALIZATION 8-1 Permit or Inhibit The balance stores parameters that must not be changed unintentionally. There are two
32Operation 1 Press the ON:OFF key to turn off the display. 2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. T
339. FUNCTION TABLE The function table reads or rewrites the parameters that are stored in the balance. These parameters stored, even if the AC adapt
349-3 Details of the Function Table Class Item and Parameter Description 0 Fast response, sensitive value FAST 1 MID.Cond Condition 2 Slow res
35Class Item and Parameter Description Cp Hi Upper limit Cp lo Lower limit Refer to “9-10 Comparator Function”. 0Key mode Accepts the PRINT key o
36Class Item and Parameter Description 0 600 bps 1 1200 bps 2 2400 bps 3 4800 bps 4 9600 bps bp5 Baud rate 5 19200 bps 0 7 bits, even
379-4 Description of the Class “Environment, Display” Condition ( Cond ) Cond 0 This parameter is for sensitive response to the fluctuation of a w
29-2 Display and Keys ...
38Zero tracking ( trc ) This function tracks zero point drift caused by changes in the environment and stabilizes the zero point. When the weight da
399-5 Description of the Item “Data Output Mode” The parameter setting of the “Data output mode ( prt )” applies to the performance when the data is
40Interval output mode The weight data is periodically output. When the PRINT key is pressed, the balance starts to output the weight data at a pre
41KF format 5if type 2 This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only communicate using this form
429-7 Data Format Examples
43 Note When “Pound Ounce” is selected, the data is output with the unit of ounce (oz). 9-8 Description of the Item “Ap
449-9 Clock and Calendar Function The balance is equipped with a clock and calendar function. When the “GLP output (info)” parameter is set to “1” o
45Confirming the time 6 The current time is displayed with all the digits blinking. When the time is correct and the date does not need to be con
469-10 Comparator Function The results of the comparison are indicated by HI OK LO on the display. Operating conditions: No comparison Com
4710. ID NUMBER AND GLP REPORT The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used. The ID number is m
31. INTRODUCTION This manual describes how the FZ-i / FX-i series balance works and how to get the most out of it in terms of performance. Read thi
48Calibration report using an internal weight When the setting is “info 1”: When the setting is “info 2”: Calibration report usi
49Calibration test report using an external weight (Calibration test does not perform calibration.) When the setting is “info 1”: When the setti
50Title block and end block When a weight value is recorded as the GLP data, “Title block” and “End block” are inserted at the beginning and at the e
5111. UNDERHOOK The underhook can be used for magnetic materials or density measurement. The built-in underhook is revealed by removing the plastic c
5212. PROGRAMMABLE-UNIT This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary coefficient set in th
5313. RS-232C SERIAL INTERFACE The balance is a Data Communication Equipment (DCE) device. Connect the balance to a personal computer (DTE) using
541 2 3 4MODEON14. CONNECTION TO PERIPHERAL EQUIPMENT 14-1 Connection to the AD-8121B Printer Set the following parameters to use the AD-8121B p
5514-2 Connection to a Computer The FX-i series balance can be connected to a personal computer using the RS-232C serial interface. As an option, t
56 Using the WinCT software, the balance can do the following: 1 Analyzing the weight data and the statistics with “RsKey” The weight data can b
5715. COMMANDS 15-1 Command List Note A command has a terminator added, that is specified using the "Terminator (Crlf)" parameter of “Seria
41-3 Compliance Compliance with FCC Rules Please note that this device generates, uses and can radiate radio frequency energy. This device has been t
5815-2 Acknowledge Code and Error Codes When the “AK, Error code (erCd)” parameter of “Serial interface ( 5if )” is set to “1”, the balance outputs
59 T command example PT command example When a communication error has occurred due to external noise, or a parity error
6016. MAINTENANCE Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is moistened with warm water and
6117. TROUBLESHOOTING 17-1 Checking the Balance Performance and Environment The balance is a precision instrument. When the operating environment or
6217-2 Error Codes Display Error code Description EC, E11 Stability error The balance can not stabilize due to an environmental proble
63 Display Error code Description EC, E00 Communications error A protocol error occurred in communications. Confirm the format, baud
6418. OPTIONS Note The FXi-02, FXi-08 and FXi-09 can not be used at the same time. FXi-02 USB interface (Installed in the balance, Applicable OS: Wi
65FXi-15-EX Carrying case You can carry the FZ-i / FX-i series balance by placing it in the carrying case. AX-FXi-31 Main unit cover Main unit
6619. SPECIFICATIONS FZ-120iFZ-200iFZ-300i FZ-1200iFZ-2000i FZ-3000iWeighing capacity 122 g 220 g 320 g 1220 g 2200 g 3200 g Maximum display
67 FX-120iFX-200iFX-300iFX-1200i FX-2000i FX-3000iWeighing capacity 122 g 220 g 320 g 1220 g 2200 g 3200 g Maximum display 122.084 g 220.084
52. UNPACKING THE BALANCE 2-1 Unpacking The balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be used
6820. EXTERNAL DIMENSIONS FZ-120i / 200i / 300i FX-120i / 200i / 300i FZ-1200i / 2000i / 3000i FX-1200i / 2000i / 3000
6921. TERMS/INDEX Terms Stable value The weight data when the stabilization indicator appears. Environment Ambient conditions such as vibration, d
70Index Keys and symbols ON/OFF key... 9 SAMPLE key... 9 MODE key ...
71Data output mode ...35 Data output pause ...35 DC static eliminator ...
72- N - NET indicator ... 9 NU format ...36, 41 Number of data ...
62-2 Installing the Balance Install the balance as follows: 1. Place the balance on a solid weighing table. Refer to “3. PRECAUTIONS” for installin
73. PRECAUTIONS To get the optimum performance from the balance and acquire accurate weight data, note the following: 3-1 Before Use Install the b